Leica Super-Resolution Microscope TCS SP8 STED

Model type
TCS SP8 STED
Manufacturer
Leica
Specification
・Live-cell confocal imaging
・Gated STED 2D super-resolution imaging with 50-nm (xy) resolutions
・5 laser lines (458, 476, 488, 514, 633 nm) and white light laser (470 – 670 nm)
・1 to 3-color STED imaging by 2 depletion laser lines (592, 660 nm)
・Motorized confocal microscope SP8 (based on Leica DMI 6000 CS)
・Adaptive Focus Control unit (maintains the desired focal plane for long-term imaging)
・Lasers Ar (458, 476, 488, 514 nm), HeNe (633 nm)
White light laser: continuous wave laser in the range of 470-670nm; AOTF allows to choose up to 8 different wavelengths simultaneously; pulsed laser system with a repetition rate of 80 MHz
Continuous wave depletion laser (592, 660 nm) for STED
Ti:Sapphire pulsed laser (Coherent Chameleon)
・Detectors
2ch PMT fluorescence detectors
2ch high-sensitive hybrid GaAsP detector HyD; time gated detection allows improvement of STED lateral (xy) resolution to <50 nm by suppressing short-lived signals at the edge of focal spot
1ch Transmitted light detector
・Scanner
FOV scanner with dual scanning galvanometer mirrors allows for either high speed scanning 7 images/sec at 512 x 512 or high pixel density scanning 8k x 8k (64 megapixels) per channel.
・Motorized XY stage and Z galvo stage for high-precision Z-sectioning
・DIC unit
・Objectives
Dry lens HC PL APO CS 10x/0.40
Oil lens HCX PL APO CS 40x/1.25-0.75, HCX PL APO CS 63x/1.40-0.60, HC PL APO 100x/1.40 STED WHITE
Water lens HCX PL APO 63x/1.20 W CORR CS
・Cell culture chamber with temperature and CO2 Control (Tokai Hit)
・Deconvolution software with STED option (Huygens)
Location
Research Bldg. No.1 Annex Rm. 301
Installation Year
2009
Check on the map
Print the map
Overview
・Live-cell confocal imaging
・Gated STED 2D super-resolution imaging with 50-nm (xy) resolutions
・5 laser lines (458, 476, 488, 514, 633 nm) and white light laser (470 – 670 nm)
・1 to 3-color STED imaging by 2 depletion laser lines (592, 660 nm)
Responsible department
Institute for Integrated Cell-Material Sciences (iCeMS)
Available Users
Kyoto University(If this is your first time, please contact us in advance.)Other Research Institutes,Companies, etc.
Notes
Usage rules
Inquiries
Institute for Integrated Cell-Material Sciences(iCeMS)
info_ac*icems.kyoto-u.ac.jp